This page provides specifications of knee specimen '''OKS010'''.  This was a pilot specimen and only imaging data is available. Courtesy of Todd Pataky and Grant 17H02151 from the Japan Society for the Promotion of Science.

<<TableOfContents>>

== Specimen Characteristics ==

Pilot knee

Demographic information not available 

'''Gender:''' '' ''

'''Age:''' 

'''Race:''' '' ''

'''Height:''' ''  ''

'''Weight:''' 

'''BMI:''' 

'''Serologically tested:''' 

''' X-ray: '''

Not available

== Experimentation ==

=== Specimen ===

'''Reference Specification:''' [[Specifications/Specimens]] - [[http://wiki.simtk.org/openknee/Specifications/Specimens?action=recall&rev=42|Revision as of FEb 4, 2019]]

'''Leading Team Member:''' Tara Nagle

'''Supporting Team Members:''' 

'''Timeline:''' 

'''Data Location:'''

Specimen characteristics are provided in this wiki page.

'''Protocol Deviations:'''

 1. X-ray evaluation was not possible due to lack of X-ray imaging of the specimen.

'''Notes:'''

 1. None.

=== Specimen Preparation ===

'''Reference Specification:'''  [[Specifications/SpecimenPreparation]] - [[http://wiki.simtk.org/openknee/Specifications/SpecimenPreparation?action=recall&rev=164|Revision as of Feb 4, 2019]]

'''Leading Team Member:''' Tara Bonner

'''Supporting Team Members:'''  Robb Colbrunn, Callan Gillespie

'''Timeline:'''



'''Data Location:'''

Data will be available as part of tibiofemoral joint mechanics data (see below).

'''Protocol Deviations:'''

 1. None.

'''Notes:'''

=== Joint Imaging ===

==== Specimen Preparation ====

'''Reference Specification:''' [[Specifications/SpecimenPreparation]] - [[http://wiki.simtk.org/openknee/Specifications/SpecimenPreparation?action=recall&rev=164|Revision as of Feb 4, 2019]]

'''Leading Team Member:''' Tara Nagle

'''Supporting Team Members:''' 

'''Timeline:''' 

'''Data Location:'''

'''Protocol Deviations:'''

 1. Patella marker was not attached.
 1. CT was also acquired for the specimen.

'''Notes:'''

==== Imaging ====

'''Reference Specification:''' [[Specifications/ExperimentationAnatomicalImaging]] - [[http://wiki.simtk.org/openknee/Specifications/ExperimentationAnatomicalImaging?action=recall&rev=121|Revision as of Feb 04, 2019]]

'''Leading Team Member:''' Tara Nagle

'''Supporting Team Members:''' 

'''Timeline:'''



'''Data Location:'''

Data was received from the imaging facilities and uploaded to in-house data management server (http://cobicore.lerner.ccf.org/midas; only accessible within Cleveland Clinic network). Raw DICOM files were stored in the folder Open Knee(s) --> Private --> oks010 --> MRI --> DICOM

DICOM files were converted to NIfTI files, stored in the folder Open Knee(s) --> Private --> oks010 --> MRI --> NIFTI of the in-house data management server. 

Data storage was accomplished based on the updated imaging specifications, see [[http://wiki.simtk.org/openknee/Specifications/ExperimentationAnatomicalImaging?action=recall&rev=121|Revision as of Feb 4, 2019]].


'''Protocol Deviations:'''

 1. None.

'''Notes:'''


 1. General purpose imaging



 1. Cartilage imaging


 1. Connective-tissue imaging



=== Joint Mechanics ===



==== Specimen Preparation ====

'''Reference Specification:''' [[Specifications/SpecimenPreparation]] - [[http://wiki.simtk.org/openknee/Specifications/SpecimenPreparation?action=recall&rev=164|Revision as of Feb 4, 2019]]

'''Leading Team Member:''' Tara Bonner

'''Supporting Team Members:''' Robb Colbrunn, Callan Gillespie

'''Timeline:'''



'''Data Location:'''

'''Protocol Deviations:'''

 1. None.

'''Notes:'''


==== Testing ====

'''Reference Specification:'''  

'''Leading Team Member:''' Tara Bonner, Robb Colbrunn

'''Supporting Team Members:''' Callan Gillespie

'''Timeline:'''

'''Data Location:'''

Data was received from the robotics testing facility and uploaded to in-house data management server (http://cobicore.lerner.ccf.org/midas; only accessible within Cleveland Clinic network). Zipped folders were stored in the folder Open Knee(s) --> Private --> oks010 --> !JointMechanics under relevant joint related folders.

Zipped folders were extracted and organized in the dissemination folder under Open Knee(s) --> Private --> oks010 --> !JointMechanics --> DISSEMINATION of the in-house data management server. 

Data storage was accomplished based on the updated joint mechanics testing specifications, see [[http://wiki.simtk.org/openknee/Specifications/ExperimentationJointMechanics?action=recall&rev=102|Revision as of Feb 4, 2019]].

'''Protocol Deviations:'''

 1. No patellofemoral joint testing is performed. 
 1. Tibiofemoral joint testing protocols were not followed as per specifications.  

'''Notes:'''

=== Tissue Mechanics ===

==== Specimen Preparation ====


'''Leading Team Member:''' 

'''Supporting Team Members:''' 

'''Timeline:''' 


'''Data Location:'''

'''Protocol Deviations:'''

 1. 

'''Notes:'''

 
==== Testing ====

'''Reference Specification:''' 

'''Leading Team Member:''' 

'''Supporting Team Members:''' 

'''Timeline:'''

'''Data Location:'''

'''Protocol Deviations:'''

 1. 

'''Notes:'''

 1. 

== Data Analysis ==

== Modeling & Simulation ==