Differences between revisions 2 and 3
Revision 2 as of 2017-05-09 19:41:34
Size: 3796
Editor: schimmt
Comment:
Revision 3 as of 2017-05-09 19:49:40
Size: 4027
Editor: schimmt
Comment:
Deletions are marked like this. Additions are marked like this.
Line 18: Line 18:
== Workflow == = Workflow =
Line 26: Line 26:
 * Double bag using a sterile technique
 * Bring USB to transfer MRI data
 * CT data is transferred via the Cleveland Clinic servers
 * Transport to MRI/CT
Line 27: Line 31:
== Position/Orient Specimen in MRI Machine ==
 * Anatomical position, anterior up. Distal end of extremity to distal end of MRI.
== Position/Orientation in MRI/CT ==
 * Anatomical position
* Anterior Up   * Distal end of extremity in first
Line 30: Line 37:
For each segment (upper leg, lower leg, upper arm, and lower arm), the following set of image collection protocols will be performed. All images should be acquired in the same coordinate system to be able to align reconstructed tissue geometries during assembly of the full segment geometry. To accomplish this the origin (isocenter) and the axes of the magnet, which is set at the beginning of the session, should not change. In addition, the specimen should not be moved. It should be noted that a pixel-by-pixel alignment of image sets (co-localization) is not necessary. For each segment (upper leg, lower leg, upper arm, and lower arm), the following set of image collection protocols will be performed. All images should be acquired in the same coordinate system to be able to align reconstructed tissue geometries during assembly of the full segment geometry. To accomplish this
 * T
he origin (isocenter) and the axes of the magnet, which is set at the beginning of the session, should not change.
 * T
he specimen should not be moved. It should be noted that a pixel-by-pixel alignment of image sets (co-localization) is not necessary.
Line 32: Line 41:
[Examples to come] Before, each sequence is acquired, verify that the acquisition properties match the desired settings in the tables below. After each sequence is acquired, inspect the images and compare them against the sample images below to to ensure that it was collected properly. If not, the sequence should be recollected.  * Before, each sequence is acquired, verify that the acquisition properties match the desired settings in the tables below. After each sequence is acquired, inspect the images and compare them against the sample images below to to ensure that it was collected properly. If not, the sequence should be recollected.
Line 34: Line 43:
= Final Imaging Specifications = == MRI Specifications ==
Line 64: Line 73:
== CT Specifications ==
 * awaiting full spec list
|| ''Slice thickness/gap (mm/mm)'' || 0.6/0 || || ||

Prerequisites Protocols

Initial Goal

A set of magnetic resonance (MR) and computer tomography (CT) images of the extremity specimen (upper/lower leg, upper/lower arm) with clear visual delineation of skin, fat, and muscle layers.

Additional information about the imaging facility, all associated contacts, and MRI hardware can be found at:

Workflow

Schedule MRI/CT

Preparation

  • Follow the instructions found here, they include registration marker attachment and placement in an imaging fixture

  • Double bag using a sterile technique
  • Bring USB to transfer MRI data
  • CT data is transferred via the Cleveland Clinic servers
  • Transport to MRI/CT

Position/Orientation in MRI/CT

  • Anatomical position
  • Anterior Up
  • Distal end of extremity in first

Acquire Image Sequences

For each segment (upper leg, lower leg, upper arm, and lower arm), the following set of image collection protocols will be performed. All images should be acquired in the same coordinate system to be able to align reconstructed tissue geometries during assembly of the full segment geometry. To accomplish this

  • The origin (isocenter) and the axes of the magnet, which is set at the beginning of the session, should not change.
  • The specimen should not be moved. It should be noted that a pixel-by-pixel alignment of image sets (co-localization) is not necessary.
  • Before, each sequence is acquired, verify that the acquisition properties match the desired settings in the tables below. After each sequence is acquired, inspect the images and compare them against the sample images below to to ensure that it was collected properly. If not, the sequence should be recollected.

MRI Specifications

Image Acquisition Properties: Download T1 SettingsDownload PD Settings

T1 w/o Fat Sup

T1 w/ Fat Sup

Proton Density

Plane

Axial

FS

No

Yes

No

Matrix (phase)

389

Matrix (freq.)

512

No. of slices

25

No. of Seq/Segment

4

FOV (mm)

240 x 202.5

Slice thickness/gap (mm/mm)

2/1

Flip angle (deg.)

30

150

TE/TR (ms/ms)

11/500

9.7/4000

Bandwidth (Hz/pixel)

238

222

Chemical shift (pixels)

N/A

No. excitations averaged

1

Turbo Factor

3

Phase encode axis

Anterior-posterior

Distance factor (%)

50%

Phase oversampling

0

Slice oversampling

0

Phase resolution (%)

90

Phase partial Fourier (8/8 = 1)

OFF

Readout partial Fourier (8/8 = 1)

OFF

Slice partial Fourier (8/8 = 1)

7/8

X-resolution (mm)

0.5

Y-resolution (mm)

0.5

Scan Time / Seq. (min.)

~3 min

CT Specifications

  • awaiting full spec list

Slice thickness/gap (mm/mm)

0.6/0

Specifications/Imaging (last edited 2019-02-06 18:30:39 by owings)